Home » Products » LAB Instruments » Clinical chemistry Instruments
Automatic analyser - Spin640 Plus

Automatic analyser - Spin640 Plus

Automatic analyser - Spin640 Plus

1/ System Function
- Fully automated, random access for routine STAT sample priority
- Throughput: 400 tests/hour, up to 640 tests/ hour with ISE
- Measuring principles: Absorbance photometry, turbidimetry
- Methodology: End-point, Fixed-time, Kinetic, optional ISE Single/Dual/Triple/Quadruple reagent chemistries, monochromatic / bichromatic
- Programming: User defined profiles and calculation

2/ Sample Handling
- Sample tray: 90 positions for primary or secondary tubes and sample cups
- Sample volume: 1.5~45 µl, step by 0.1 µl
- Sample probe: Liquid level detection, clot detection and collision protection
- Probe cleaning: Automatic washing both interior and exterior, carry-over ≤ 0.05%
- Automatic sample dilution: 
   + Pre-dilution and post-dilution
   + Dilution with ratio up to 1: 150
   + Dilution vessel: Cuvette

3/ Internal Bar Code Reader (optional)
Used for sample and reagent programming; applicable to various bar code systems including Codabar, ITF (Interleaved Two of five), code128, code39, UPC/ EAN, Code93; capable to communicate with LIS in a bi-directional mode

4/ ISE Module (optional)
- Optional selection of K+, Na+, Cl-, or K+, Na+, Cl-, Li+
- Throughput: Up to 240 tests per hour

5/ Reagent Handling
- Reagent tray: 80 positions in refrigerated compartment (2~10°C)
- Reagent volume from 10µl to 350µl; 2 independent reagent probes
- Reagent probe: Liquid level detection, collision protection and inventory checking
- Probe cleaning: Automatic washing both interior and exterior

6/ Reaction System
- Reaction rotor: Rotating tray, 90 cuvettes with automatic washing
- Cuvette: Optical length 5mm
- Reaction volume: 150~360 µl
- Operating temperature: 37°C
- Temperature fluctuation: ±0.1°C
- Mixing system: 2 independent mixers

7/ Optical System
- Light Source: Halogen-tungsten lamp
- Photometer: Reversed optics, grating photometry
- Wavelength: 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm, 800nm
- Absorbance range: 0~3Abs (10mm conversion)
- Resolution: 0.001Abs

8/ Control and Calibration
- Calibration mode: Linear (one-point, twopoint and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, exponential, Polynomial, Parabola
- Control rules: Westgard multi-rule, Cumulative sum check, twin plot

9/ Operation Unit
- Operation system: Windows® XP Professional/Home SP2 Windows® Vista Home/Business
- Interface: RS-232

10/ Working Conditions
- Power Supply: 200~240V, 50Hz, 1500VAor 220/230V, 60Hz, 1500VAor 110/115V, 60Hz, 1500VA
- Temperature: 15~30°C
- Humidity: 35~80%
- Water consumption: 20L/hour
- Dimension: 1180mm x 700mm x 1145mm (W x D x H)
- Weight: 300 Kg